Disposal
Disposal of the device via public waste disposal systems is not permissible. Contact
your local MACHEREY‑NAGEL contact person.
Disposal according to EU directive 2012/19/EU. In accordance with EU directive
2012/19/EU, MACHEREY‑NAGEL takes back old equipment and disposes of it free of
charge.
Troubleshooting
Various messages may be displayed, depending on the operating status. The source of the
error may either be incorrect operation or a malfunction of the device. If errors recur, contact
MACHERY‑NAGEL.
Problem
The barcode could not be read.
The cuvette used is the wrong size.
Device is not ready for
measurements.
The wavelength is outside of the
permissible range.
You are not authorized to perform
this action.
The export failed.
Quickstart Guide NANOCOLOR
20
NOTICE
Cause
The cuvette inserted
does not have a
barcode.
The barcode on the
cuvette is damaged.
The device does not
recognize the test kit
used.
The cuvette size
programmed for the
method does not
match the cuvette size
determined.
Automatic device tests
display errors.
The wavelength entered
is outside of the
wavelength range of the
device.
The user is not
authorized to change
this setting.
Printing on a USB drive
was not possible.
/
® UV
VIS
Solution
Call up the respective test manually and insert
the cuvette according to the instrument's
instructions.
Manually enter the test number of the method
used.
Update your instrument to the latest firmware
version (see operating instructions, chapter
6.3.).
Check whether the correct cuvette size is
used for the applied method.
Check whether the cuvette was correctly
inserted.
Restart device. If the error persists, copy
the log file onto a USB drive (see operating
instructions, chapter 6.2) and contact your
local distributor or MACHEREY‑NAGEL.
Enter a wavelength in the correct range (see
operating instructions chapter 1.5).
Change the user rights or contact the device
administrator to change the setting.
Check whether write protection of the USB
drive is disabled.
Try the export with a different USB drive.
Try the export with a different USB‑A interface.
II / VIS II V 2.00/09‑2022