Designation
Article number
Dimensions L x W x D in cm
Weight in kg
Material composition
Location of type plate
3.3 EMC conformity
The BEMER therapy system Evo is intended for operation in an electromagnetic environment in
which immunity to RF disturbances is uncontrolled.
Electromagnetic Emission
Emission
Measurements
Conformity
RF_emissions in
Group 1
accordance with CISPR 11
RF_emissions in
Class B
accordance with CISPR 11
Harmonics in accordance
Class A
with IEC 61000-3-2
Voltage fluctuations/flicker
Compliant
in accordance with IEC
61000-3-3
B.Body Evo foot protection
450500
67.5 x 60.0 x 0.02
0.23
PES
Product packaging
Electromagnetic Environment –
Guidlines
The BEMER therapy system uses RF energy
only for its internal function. As a result,
its RF emission is very low and it is unlikely
to disturb any adjacent electronic devices.
The BEMER therapy system is intended for
use in all institutions including residential
areas and those that are directly connected
to a public grid that also supplies buildings
used for residential purposes.
32
Recommended protective distances between portable and mobile RF telecommunication
devices and the BEMER therapy system
Customers or users can help to avoid electromagnetic disturbances by complying with the minimum
distance between portable and mobile RF telecommunication devices (transmitters) and the BEMER
therapy system, depending on the output of the communication device as indicated in the table on
the next page.
Rated output of
Protective distance,
transmitter [W]
depending on transmission frequency [m]
150 kHz to 80
MHz outside
ISM bands
3.5
d=
√P=1.17√P
3
0,01
0,12
0,1
0,37
1
1,17
10
3,69
100
11,67
For transmitters where the maximum rated output is not listed in the table above, the distance can
be determined by using the equation that belongs to the relevant column, in which P is the maximum
rated output of the transmitter in watts (W) according to information from the transmitter manufacturer.
ES
NL
IT
150 kHz to 80
80 MHz to
800 MHz to
MHz inside
800 MHz
2,5 GHz
ISM bands
12
12
23
d=
√P=1.2√P
d=
√P=1.2√P
d=
10
10
10
0,12
0,12
0,23
0,38
0,38
0,73
1,2
1,2
2,3
3,79
3,79
7,27
12
12
23
33
FR
DE
EN
√P=2.3√P