Angled Screwdriver ANGULUS
3.3
Sterilization
WARNING
Danger of infection due to non-sterile handling!
Improper sterilization and non-sterile handling of the angled screwdriver poses serious health
hazards to patients.
Sterilization must be carried out according to a validated steam sterilization process, for
example in a sterilizer satisfying EN 285:2009, ANSI / AAMI ST 79 requirements and
validated in accordance with ISO 17665-1:2006 requirements.
ANSI / AAMI ST 79 recommends a minimum cycle time of 4 minutes at 134 °C (269.6 °F) for
dynamic air removal steam sterilization cycles. Please follow the instructions of the user
manual of your steam sterilizer.
NOTICE
The responsibility for proper cleaning, disinfection and re-sterilization rests with the operator /
product user. Be sure to observe your national / local regulations, including potential
restrictions.
Use appropriate and approved sterile packaging (e.g. in accordance with EN 868, ISO 11607)
for sterilization, subsequent transport and storage.
V. 2.0
The shank (2) without gear unit, is also
sprayed from the inside.
Insert the INTRA coupling adapter with the
reduction piece into the shank and flush the
shank until clean oil is coming out at the
other end.
Remove any excess oil manually with a soft,
lint-free and absorbent cloth.
33