WARNING
Danger of injury from aspiration and swallowing of small parts!
The angled screwdriver consists of a multitude of components and accessories. Be sure to
check the instrument for completeness and integrity (particularly the drill bit) each time you
remove it from the oral cavity. Parts that may have come off or fractured off (e.g. as a result
of bit fracture) must be removed from the oral cavity, the throat and / or the bone at once.
CAUTION
Danger of burns from hot components (drive, shank, drill bit)!
The angled screwdriver may be operated at a maximum speed of 5,000 rpm (revolutions per
minute).
Using higher speeds, as well as using the product in a non-intended way, can heat up the
shank and the drive components to more than 40 °C (104 °F).
To prevent bone necroses, drill bits must be sufficiently cooled.
36
Position the screw in the drilled hole.
•
To insert the screw, rotate the handle
•
part (4) clockwise.
During this process, retract the screw holder
•
mechanism as soon as the screw takes hold
in the bone.
Instructions for Use
Angled Screwdriver ANGULUS
V. 2.0