To sterilize the components of the angled screwdriver and its accessories, we recommend
storing them in the optionally available mesh tray and small parts tray (for pictures see
section 3.2 "Machine Cleaning and Disinfection", page 31) and using a suitable sterilization
container (e.g. KLS Martin Micro Stop
3.4
Reprocessing limitations, Disposal
Frequent reprocessing has only a minor adverse effect on the angled screwdriver and / or its
components. The life of the product is mainly determined by wear and potential damage during
use.
Reprocessing the angled screwdriver and its components is limited to 100 processing cycles.
Once the instrument has reached the end of its useful life, it must be properly disposed of
•
or recycled.
4
Inspection Prior to Reuse
Before every application of the instrument, check it for fractures, cracks, deformation,
•
damage and proper function.
Worn, corroded, deformed, porous or otherwise damaged instruments must be disposed
•
of.
34
®