5.4 Correction via Link2sc
The Link2sc procedure offers a secure method of data exchange between process probes and
Link2sc-compatible photometers using an SD memory card or via a local area network (LAN).
During a pure control measurement, the measurement data is transferred from the probe to the
photometer where it is then archived together with the photometric reference data that has been
recorded.
Refer to the Link2sc user manual for a detailed description of the Link2sc procedure.
5.5 Data logging
The SC Controller provides one data log and one event log for each instrument. The data log stores
the measurement data at selected intervals. The event log stores different types of events that occur
on the devices (i.e., configuration changes, alarms and warning conditions).
The data log and the event log can be saved. Refer to the SC Controller user manual for instructions.
Section 6 Maintenance
Multiple hazards. Only qualified personnel must conduct the tasks described in this section of the
document.
Do not disassemble the instrument for maintenance. If the internal components must be cleaned or repaired,
contact the manufacturer.
6.1 Maintenance schedule
Table 3
shows the recommended schedule of maintenance tasks. Facility requirements and
operating conditions may increase the frequency of some tasks.
life of wearing parts for standard operating conditions and default (factory) settings.
Task
Visual inspection
Clean the measuring path
Validate the probe calibration
Replace the wiper blade
Manufacturer service inspection
Wearing part
Wiper blades (set of 5 pieces)
Wiper motor
Wiper shaft with seals
10
Based on non-abrasive water conditions
11
Based on the default settings
12
A minimum of one inspection per year is required. For the best performance and uptime, the
manufacturer recommends 2 inspections per year.
N O T I C E
Table 3 Maintenance schedule
on page 18
on page 20
on page 19
Table 4 Wearing parts consumption
C A U T I O N
Table 4
1 week
3 months
X
10,11
X
Quantity
1
1
1
shows the average service
6 months
1 year
As necessary
10
X
12
X
Average service life
10,11
> 1 year
11
7 years
11
2 years
X
English 17