11 Additional functions
1. Calculating the adjustment value
The already adjusted instrument shows a new deviation of the mean volume from the nominal vol-
ume (e.g., 0.017 ml). For information on calculating the mean volume, see Checking the Volume (Cali-
bration), p. 61.
2. Entering CAL mode
a.
b.
3. Entering the adjustment value
a.
64
Operating manual
a.
With the instrument powered on, press and hold the
> 3 s
CLEAR button for more than 3 seconds until the follow-
ing modes are repeatedly shown one after another in
the display:
CAL — GLP — APO — dP
b.
When CAL appears in the display, release the CLEAR
button. CAL blinks and the adjustment value of the pre-
vious adjustment appears.
a.
The adjustment value is 0.017 ml, for example. Press the
Pause or On/Off button until the value has been
reached (the first press of a button sets the display to
zero).
-
+