4. Confirming input
a.
In rare cases, when entering the new adjustment value, the sum of the old and new adjustment values
can amount to zero. In this case, the factory setting is restored and CAL disappears from the display.
11.1.3 Factory default setting
The continuously displayed CAL symbol indicates that an adjustment has already taken place. How-
ever, the factory default setting should be restored.
1. Entering CAL mode
a.
b.
997469
a.
Press the CLEAR button. The old and new adjustment
values are added automatically.
ð
The change in adjustment is indicated by the CAL sym-
bol.
NOTICE
a.
With the instrument powered on, press and hold the
> 3 s
CLEAR button for more than 3 seconds until the follow-
ing modes are repeatedly shown one after another in
the display:
CAL — GLP — APO — dP
b.
When CAL appears in the display, release the CLEAR
button. CAL blinks and the adjustment value of the pre-
vious adjustment appears.
11 Additional functions
65
Operating manual