Instructions for Use
The IDP-10 is designed and certified for operation indoors (installation category II)
in relatively clean, dry environments (pollution degree 2).
Refer to Table 15 for specifications. It is not designed nor certified for operation
outdoors.
The IDP-10 should never be used in ambient environments that contain explosive or
combustible materials. The IDP-10 is NOT intrinsically safe and should never be
used with corrosive, pyrophoric, particulate-forming gases or gas mixtures in their
flammable and/or combustible range.
Connect the pump to the power supply using an IEC-320 style power cord.
During operation the following environmental conditions must be respected:
CAUTION
•
•
Connect the pump to the power supply using an IEC-320 style power cord of at
least 10 A capacity (See Table 16).
When installing the vacuum pump it is recommended that the pump be
CAUTION
positioned in a way that the power switch and power cord connection remain
easily accessible to the user in order to be available as an option to disconnect
power to the pump quickly if needed.
Agilent IDP-10 Dry Scroll Vacuum Pump User Manual
Temperature: +5 °C to +40 °C (41 °F to 104°F)
Relative humidity:0 to 95% (non-condensing)
171