Descargar Imprimir esta página

Agilent IDP-10 Manual Del Usuario página 185

Publicidad

Technical Information
Technical Specifications
Table 15 Technical Specifications
Model
Interface dimensions
Peak pumping speed
Media
Ultimate pressure*
Maximum recommended continuous inlet
pressure
Maximum inlet pressure at nominal voltage
Maximum outlet pressure
Inlet connection
Exhaust connection
Gas ballast
Ambient operating temperature
Storage temperature
Motor rating
Operating voltages**
Motor thermal protection
Operating speed
Cooling system
Weight
Shipping weight: 28.1 kg (62 lbs)
Agilent IDP-10 Dry Scroll Vacuum Pump User Manual
IDP-10 Dry Scroll Single Hermetic Vacuum Pump
H=297 mm x W=260 mm x L= 420mm
3
170L/min (10.2 m
/h)
No corrosive, explosive or particulate forming
gases
1.5 x 10-2 Torr (2.0 x 10-2 mbar)
300 Torr (400 mbar)
1.0 atmosphere (0 psig)
115V range: 5 psig, 115 VAC 50/60HZ or greater,
derate linearly to 0 psig at 100VAC below 115VAC
230V range: 5 pisg, 200 VAC 50/60HZ or greater
NW25
Female 1/4" National Pipe Thread (NW16)
Female 1/8" National Pipe Thread, (20 micron
sintered plug provided)
5 °C to 40 °C (41 °F to 104 °F)
–20 °C to 60 °C (–4 °F to 140 °F)
1/2 HP (370 W)
Operates with inverter - functions with several
power inputs: 1 phase/100-120VAC 50/60Hz, 9 A
typical; 200-240 50/60Hz, 5 A typical
Automatic thermal protection
Full speed is 1900 RPM; RPM can be adjusted
between 1350 to 1900 RPM.
Factory setting for X3810-64000 and X3810-64010
is 1900 rpm, there is optional default setting at
1800 rpm available.
Air-cooled
Pump only: 24.7 kg (55 lbs)
185

Publicidad

loading

Productos relacionados para Agilent IDP-10