Measuring AC voltage
The AC voltage measurement function has the following features:
– Five ranges to select: 100 mVrms, 1 Vrms, 10 Vrms, 100 Vrms, and 250 Vrms,
– Measures the AC-coupled true RMS.
– Measures within the stipulated accuracy at a crest factor of a maximum of 5:1
– Input impedance is 1 MΩ ± 2% in parallel with less than 120 pF capacitance on
Make the connection as shown below.
Keysight Measurement Manager operation
Select the ACV function and desired range. A suitable range should be selected to
give the best measurement resolution. The reading is displayed and updated
continuously.
SCPI commands
The following example shows how to make an AC voltage measurement using
SCPI commands.
MEASure[:VOLTage]:AC?