Point-of-care studies
Point-of-care studies have been performed at three individual district
health centers in primary care with both nursing and physician services.
The studies were performed on capillary samples in duplicates tested
on HemoCue Hb 301 vs HemoCue Hb 201
against the International Council for Standardization in Haematology
method (ICSH)
3
using venous blood (2).
Study
N
Min
Max
Regression
g/dL
g/dL
line
1
186
8.3
18.0
Y = 0.897x +
1.357
2
700
1.6
23.0
Y = 1.005x
-0.041
Technical Specifications
Dimensions: 160 x 140 x 70 mm (6.29 x 5.51 x 2.76 inches)
Weight: 500 g (1.10 pounds) (with batteries installed)
4 batteries type AA
Power adapter: CE marked
Only use adapters, as listed under "Adapters".
Pollution degree: 2
Overvoltage category: II
Atmospheric pressure: 700 hPa to 1060 hPa.
Equipment not suitable for use in the presence of flammable mixtures.
The instrument is tested according to
IEC 61010-1, Second edition: 2001 and EN 61010-1: 2001, IEC/
EN 61010-2-101: 2002, IEC 61326-1: 2012 and EN 61326-1: 2013,
IEC 61326-2-6: 2012 and EN 61326-2-6: 2013, IEC 60601-1-2, Third
edition: 2007 and EN 60601-1-2: 2007 and complies with the IVD Medical
Device Directive 98/79/EC.
The instrument is made for continuos mode.
Essential Performance
The essential performance is quantitative determination of hemoglobin
in capillary, venous or arterial whole blood.
AC Adapters
(1). The Hb 201
was verified
+
+
Region: EU/US/UK/Universal
Type: 12 V HCA01
Input: 100 V~-240 V~/50-60 Hz/≤250 mA
Correlation
coefficient (r)
0.930
0.999
46