Technical specifications (EMC-RF)
Guidance and manufacturer´s declaration – electromagnetic emissions
The HemoCue systems are intended for use in the electromagnetic environment specified below. The customer or the user of the HemoCue systems
should assure that it is used in such an environment.
Emission test
RF emissions
RF emissions
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/flicker emissions
IEC 61000-3-3
NOTE 1
It is the manufacturer's responsibility to provide equipment electromagnetic compatibility information to the customer or user.
NOTE 2
It is the user's responsibility to ensure that a compatible electromagnetic environment for the equipment can be maintained in order that
the device will perform as intended.
Use only cables with the following specification:
USB shielded maximum 2 m
Serial shielded maximum 1.5 m
Compliance
Group 1
Class B
Class A
Complies
51
Electromagnetic environment – guidance
The HemoCue systems uses RF energy only for its internal
function. Therefore, its RF emissions are very low and
not likely to cause any interference in nearby electronic
equipment.
The HemoCue systems are suitable for use in all
establishments, including domestic establishments and
those directly connected to the public low-voltage power
supply network that supplies buildings used for domestic
purposes.