5.3
USING THE SOLAR-02 IN INDEPENDENT MODE
1. Turn on the instrument by pressing the /ESC key.
2. Press the SET key. The unit shows the screen beside in which the K value of
the currently set sensor is selected.
3. Use the arrow keys ▲and ▼ to select the desired type of sensor considering
the unit programming (see § 0)
4. Confirm selection with the SET key to go back to the measuring mode
5. Position the irradiation measuring probe type "Irr-P" or "Irr-S2" (see Table 1)
on the modules' surface and connect the relevant output lead to the PYRA/
CELL input of the unit SOLAR-02
6. Fasten the temperature probe type Tmp with adhesive tape (see Table 1) on
the rear side of the modules and connect the relevant output lead to the
TEMP input of the unit SOLAR-02
7. Read on the display, in real time, as shown in the figure beside, the values of:
• Irradiation on the main display, expressed in W/m²
• Tc = temperature of cells expressed in °C
• Te = temperature of environment expressed in °C
5.4
USING THE INCLINOMETER FUNCTION
1. Turn on the unit by pressing the /ESC key. The display shows the following
screen
2. Press the FUNC/▲key to activate the inclinometer mode of the unit
SOLAR-02 instead of the measurement of temperature Te as shown on the
screen beside
3. Fasten the unit on the module's surface, if necessary by means of the two
magnets available on the rear side
4. Wait for a few seconds
5. Read, in real time, the value (expressed in °) of the tilting angle against the
horizontal plane of the object to be tested
"
5.5
USING THE SOLAR-02 WITH MASTER INSTRUMENT TYPE "M-USB"
"M-USB"
For the definitions of the instrument types, see Table 1. For the use, functions
and activation of a testing/recording on the master instrument, please see
the user manual of the master instrument itself.
5.5.1
PRELIMINARY CHECK
For the "M-USB" master instruments the enable of internal RF module is NOT
necessary. Check that the parameter which enable/disable the RF module
should be corrected set at the first usage of the unit in the herewith way:
By press and hold the FUNC/▲key while switching on the meter by
key. The message "rF
" is displayed for a few seconds. If the message "rF
OFF
is displayed switch off the unit and repeat the operation. Afterwards the
initial screen is displayed by the meter. The setting of "RF" parameter is
maintained also during the replace of batteries