3.4 Measurement
A measurement is started by selecting the option Measure in the main
menu and following the instructions that appear on the display. A
measurement always consists of:
1. Sample preparation
2. Sampling
3. Analysis
4. Result
5. Regeneration
Sample preparation
Always prepare samples freshly right before measurements.
Your sample liquid, the vials and pipettes must be at ambient
temperature to avoid measurement errors.
Reusing sample vials or pipettes will lead to wrong results.
1. Withdraw 2 mL of your sample liquid with a fresh
pipette.
2. Release the liquid into the vial and onto the filler
material.
CAUTION
Do not wet the windings or rim of the vial. This will lead
to inaccurate result and can poison the device.
If the device comes into contact with sample liquid, stop
the measurement, clean the areas with a wet cloth and
let the device dry before the next measurement.