Check if the sealing O-ring (see chapter 2.1) is missing. If so,
replace it with the included replacement or otherwise contact
Alivion's technical support.
Check if the windings are damaged in any way. If so, please
contact Alivion's technical support.
Sampling failed (ERR 942)
Do not remove the sample vial before prompted to do so.
Ensure that the sample vial is not moved or tilted during
sampling.
Make sure the vial is screwed firmly and try using a new vial.
Make sure the Sealing O-Ring is installed correctly.
Contact Alivion's technical support if the problem persists.
Sampling failed (ERR 943)
Ensure that exactly 2 mL of sample is poured into the sample
vial.
Ensure the adsorbent material in the sample vial is neither
compressed nor removed.
Make sure you use an original Alivion sample vial.
7.2 Measurement errors
Measurement failed (ERR 950/951)
Ensure the inlet of the Alivion Spark M-20 is clean. Remove
debris, liquid, or alcoholic sample at the inlet of the device.
Do not screw in a sample vial or block the inlet in any other way
during the analysis.
Measurement failed (ERR 960/961)
Retry the measurement with a freshly prepared sample.
Retry the measurement with a different sample liquid.
Try to clean the sensor (Settings → Resets → Clean sensor).
Contact Alivion's technical support if the problem persists
Alivion Spark M-20
Troubleshooting
19