Troubleshooting
Critical error (952/953)
Restart the device and retry the measurement.
Contact Alivion's technical support if the problem persists.
7.3 Calibration errors
Calibration failed. Measurement error
The calibration failed due to a measurement error described in
chapter 7.2. Follow the instruction for the respecting
measurement error.
Temperature difference during calibration too large
Ensure the ambient temperature does not change during the
first and second calibration measurement. Retry the calibration
once the temperature is constant.
Calibration failed (ERR 972 -975)
Ensure the correct standard was used for the calibration.
For a 1-point calibration Alivion Methanol Standard 1 (orange) is
used. For a 2-point calibration, Alivion Methanol Standard 1
(orange) is used first, followed by Alivion Methanol Standard 2
(blue).
Ensure only original Alivion calibration standards and sample
vials are used.
Ensure the inlet of the device is clean and free of debris, liquid,
or alcoholic sample.
Ensure any used sample vials and pipettes are not near the
inlet of the Spark M-20.
Ensure no sample liquid is on the windings or rim of the sample
vials.
Try to clean the sensor (Settings → Resets → Clean sensor).
20
Alivion Spark M-20