3 Use
The device is delivered with a factory calibration. However,
Alivion strongly recommends performing a 2-point calibration
before the first measurement in order to achieve the maximum
measurement accuracy.
The device must be calibrated regularly. Alivion recommends
performing a new calibration at the beginning of each
measurement day in order to guarantee maximum accuracy.
The device can be used for measurements within the
operational temperature range of 10–35 °C and delivers
accurate results using temperature correction algorithms.
While the device accounts for temperature variations, the
measurement is highly temperature sensitive. Please carefully
follow the instructions in chapter 3.4 to achieve maximum
accuracy.
3.1 Recharging the battery
Charge the device by connecting it to a wall socket using the provided
Alivion USB charger.
The device battery level is indicated with the corresponding icon in the
top right part of the screen.
Fully charged
2
3 ⁄ full
1
3 ⁄ full
Almost empty
The Li-ion polymer battery can be charged at any time.
Interrupting the charging process will not damage the battery.
Avoid keeping the battery at 100% all the time, as this may
cause accelerated aging. Therefore, it is recommended to
charge the battery if the charge level is 1 3 ⁄ or less.
Alivion Spark M-20
NOTICE
Measurements are not
possible during charging.
Use
7