6.3 Change seal
Remove old seal
b.
c.
1
2
Check instrument
a. Check the volume after each change of capillary, piston rod, or seal.
See also
→
Ordering data for instruments and spare parts (p. 50)
997916
a. Remove capillary (see Change capil-
lary, p. 40 ).
b. Cut the old seal with a sharp knife
(scalpel or similar), and remove it.
c. Insert a new seal (1) into the attach-
ment block (2) (included in the repair
set, see page 22).
d. Press the piston rod firmly into the
seal.
e. Pull off the attachment block.
f. Check that the seal is firmly seated.
g. Slide on the capillary (see Change
capillary, p. 40 ).
Operating manual