9 Accuracy table
Volume
Accuracy R ≤ ±
%
Variable
2.5–10
3.0
5–25
2.4
10–50
1.8
20–100
1.5
Fixed
1
12.0
2
7.5
5
3.0
10
3.0
20
2.4
25
2.4
50
1.8
100
1.5
200
1.5
20 °C
!
Ex
Final test values based on the nominal volume (= max. volume) printed on
the device and the specified partial volumes at the same temperature (20
°C/68 °F) of the device, surroundings and distilled water, in accordance
with DIN EN ISO 8655.
997916
µl
0.3
0.6
0.9
1.5
0.12
0.15
0.15
0.3
0.48
0.6
0.9
1.5
3
Coefficient of variation CV ≤
%
µl
0.8
0.08
0.5
0.125
0.4
0.2
0.4
0.4
4.0
0.04
2.0
0.04
0.8
0.04
0.8
0.08
0.5
0.1
0.4
0.1
0.4
0.2
0.4
0.4
0.2
0.4