Disconnecting the transmitter from the sensor
Before proceeding, have your MiniMed insulin pump system user guide available.
To disconnect the transmitter from the sensor:
1. Carefully remove any tape from the transmitter and sensor.
2. Remove the adhesive tab from the top of the transmitter.
3. Hold the transmitter as shown, and pinch the flexible side arms of the sensor between your
thumb and forefinger.
4. Gently pull the transmitter away from the sensor.
5. Follow the instructions that appear on the pump or in your system user guide.
Removing the sensor
Always refer to the sensor user guide for instructions on how to remove the sensor.
Reconnecting the transmitter to a sensor that is already inserted
You can reconnect your transmitter to the sensor you are currently using. Simply connect your
transmitter to the sensor that is already inserted. When the pump detects the transmitter, confirm that
you want to Reconnect Sensor. It may take a few seconds to establish a connection when reconnecting
a sensor. Reattach the adhesive tab of the sensor to the transmitter and reapply any required tape.
When you reconnect a sensor, the sensor will go through another warm-up period before you can
calibrate it.
Tester
The tester is used to test the transmitter to make sure it is working. The tester is also used as a required
component to create a waterproof seal when cleaning the transmitter. Properly connecting the tester