Descargar Imprimir esta página

Medtronic GUARDIAN LINK 3 Manual Del Usuario página 30

Ocultar thumbs Ver también para GUARDIAN LINK 3:

Publicidad

9. Rinse the transmitter under running room temperature tap water for at least one minute, and
until all visible liquid soap is gone.
10. Dry the transmitter and tester with a clean, dry cloth.
11. Prepare a 1:10 bleach solution by using one (1) part 8.25% bleach to nine (9) parts water, for a final
concentration of 0.8%. Make sure to prepare a fresh solution for each use.
12. Make sure that you have completed the previous cleaning steps before disinfection. With the
tester still attached, soak the transmitter in the bleach solution for 20 minutes.
30

Publicidad

loading

Este manual también es adecuado para:

Mmt-7910