11. Disconnect the tester from the transmitter by gently squeezing the arms of the tester.
For multiple-patient use
When using the transmitter in a healthcare facility, always clean and disinfect the transmitter after each
use.
WARNING: You must adhere to Standard Precautions when handling or using this device. All
parts of the system should be considered potentially infectious and are capable of
transmitting blood-borne pathogens between patients and healthcare professionals.
The transmitter must be disinfected after use on each patient. This system may only be used
for testing multiple patients when Standard Precautions and disinfection procedures
provided by Medtronic Diabetes are followed.
To clean the transmitter, you need the following materials:
• gloves
• mild liquid soap
• soft-bristled toddler toothbrush
• 8.25% bleach
• two containers
• clean and lint-free dry cloths
Use life
The transmitter can be cleaned and disinfected up to 122 times or for one year, whichever comes first.
Discard the transmitter at this point. If you continue to use the transmitter beyond 122 times or one
year, the cleaning and disinfection process may damage the device. Contact Medtronic to order a new
transmitter.
27