Operation
Resetting the display prior to direction changes
Prior to a direction change, e.g. clockwise/counter clockwise, the
DPG must be reset to zero. This increases the accuracy of the test.
To reset the display in the event of direction changes, proceed as
follows:
➢ Apply the DPG's rated torque once in the test direction.
➢ Relieve the DPG.
➢ Switch the DPG off.
➢ Switch the DPG on.
The DPG is tared and is available for tests in the desired direction.
Overload warning
If a torque which exceeds the DPG's measuring range is applied, it
can lead to material damage to the DPG. The DPG is therefore
equipped with an overload warning function.
If the nominal value is exceeded by more than 10 %, the red LED
lights up and an acoustic signal is additionally sounded.
➢ Immediately abort the test procedure.
➢ To perform a new test procedure, press the "CLEAR HOLD
POWER" button.