Using the TORKMASTER 4 programme
Selection of the calibration method
One of the following test methods can be selected:
• Quick test:
The tool is tested without creating a standard-
conforming log.
• Test and adjustment:
A tool can be tested and adjusted in the case of
deviating values.
• According to EN ISO 6789:
Calibration of a tool (calibration object) according
to the standard's specifications. Results cannot be
deleted.
• Based on EN ISO 6789:
Calibration of a tool (calibration object) according
to the standard's specifications. Individual results
can be deleted and the measurement can be
repeated.
Further calibration methods are possible on
request.
A selection can only be made in this field if you
have selected a calibration object.
Serial number of the connected transducer
("Reference")
The serial number of the connected transducer is
displayed here.
In certain cases, the measuring range of the
transducer does not cover the measuring range of
the calibration object completely. In such cases, the
transducer must be changed during calibration. The
information on the two transducers is displayed and
stored in the log.
30