Maintenance
Maintenance
Maintenance work is exclusively performed on the
outside of the calibration and adjustment facility.
Proceed as follows before carrying out maintenance
work:
Switch the calibration and adjustment facility off
using the switch on the IEC power connector
socket.
Additionally remove the mains plug from the
mains socket.
Make sure that the calibration and adjustment
facility cannot be switched on without
authorisation.
Carry out the necessary maintenance work.
Check the function of the available protective
facilities.
Cleaning
ATTENTION
Malfunctions due to ingressing
cleaning agents.
Use cleaning agents sparingly to
avoid the ingress of cleaning agents
into the electrical installation space.
Carry cleaning and care work out as required
according to the following working rules:
• Do not use compressed air for cleaning the
system. Ingressed dirt damages seals, bearings
and electrical components.
• Clean all machine components, covers and
splash panels with a cloth soaked in mild soap
solution.
• Clean lines and plastic parts using a mild soap
solution.
• Observe and comply with the manufacturers'
instructions regarding cleaning agents.
• Remove escaped lubricants using a mild soap
solution.
36
• Prevent the ingress of soap solution into bearings,
seals and electrical systems.
• Clean corroded areas and, if permissible, coat
with paint, grease or oil.
• Replace damaged, illegible or missing labels and
signs.
• Store cleaning agents and lubricants according to
the applicable regulations.
• Dispose of cleaning agents and lubricants
according to the applicable regulations.
Exchanging fuses
The IEC power connector socket is fused with two
type T6.3A/250 V glass tube fuses.
DANGER
Lethal hazard due to electric shock.
Switch the calibration and
adjustment facility off using the
switch on the IEC power connector
socket.
Remove the mains plug from the
mains socket.
Proceed as follows to replace a defective glass tube
fuse:
Switch the calibration and adjustment facility off
using the switch (2) on the IEC power connector
socket.
Remove the mains plug from the mains socket.
Remove the fuse holder (1).