5.2.3. Step Voltage Test (SVT)
In this mode of operation, the user does not set a specific test voltage, but
rather the values of maximum voltage, voltage step and step time
value. The instrument will start the test by applying the value of the
voltage step and increase this value at each programmed time interval
until it reaches the maximum voltage value. At each stage, the equipment
measures the resistance before moving to the next step.
step
voltage
The test result is calculated using the following formula:
• Access the device's MENU using the
• Use the control
• Use the control
step time
R
SVT
=
R
, to select SETUP TEST and press
MODE SELECTION
SETUP TEST
LANGUAGE
, select STEP VOLTAGE TEST and press
=
SETUP TEST========
MANUAL TEST
TEMPO
RIZED TEST
STEP VOLTAGE TEST
21