• Press
to start the test. The equipment will start applying high
voltage and the high voltage LED (
test number will be displayed and then the measured resistance value:
Measured resistance
Applied voltage
Voltage selected
• Wait for the test to finish. The results screen will appear:
Measured resistance
Applied voltage
Capacitance
The capacitance value may take a few seconds to be calculated
6.5. "Pass / Fail" Test mode
The "pass / fail" test can be performed in any of the measuring modes,
simply set the MINIMUM RESISTANCE value of the selected mode (see
5.2.5.1, pg. 24). During this type of test, the equipment will indicate with
an intermittent "BEEP" when the insulation resistance is lower than the
programmed threshold.
R= 5.14GΩ
Ua:
505 V
97.9nA
Ut:
500 V
00:17
41
) will light up. The
Leakage current
Elapsed time
Elapsed time
Leakage current