Using the TORKMASTER 4 programme
Number
Designation
1
Serial number of the calibration facility and selection field for manual selection of the
calibration facility
2
Input or selection of the calibration object ("Calibration object")
3
Selection of the calibration method
4
Result of the current measurement
5
Numerical display of the current measured value
6
Bar display of the measured value in the permissible measuring range
7
Graphical portrayal of the current measured value
8
Enlarged bar display of the target value
9
Status line for messages:
• Green: Notes and information on the measurement
• Yellow: Calibration instructions
• Red: Error messages
10
Button for opening the comment field
11
Buttons for individual functions
12
Displays for the two transducers ("Reference")
Selection of the calibration facility
In the case of type 7794-2 and 7794-3 calibration
and adjustment facilities, this information is
automatically entered as soon as a calibration and
adjustment facility is connected to the computer.
This information has to be entered manually in the
case of other types.
To do this, click into the selection field.
Then click onto the entry for the connected
calibration facility.
20
Input field for the calibration object
The following information can be input in this field:
• Serial number or
• Identification number
The search criterion can be switched over in the
basic settings.
If the number is displayed in red, the calibration
object has not yet been created in the database.