1.4. INSTALLATION AND COMMISSIONING

1.4.1. INSTALLATION

Follow these steps to assemble the booth:
Choose the location where the booth is to be placed and its layout. Decide where
the front panel is to be located, which includes the window and the table, and
on what side of the front panel the door will be hung, as well as its opening
direction.
Note: After selecting the location of the front panel, the rear panel will go on the
opposite side; after choosing the location of the door, the side panel should go
on the opposite side.
Check the assembly guide (501-708-GR_) supplied with the booth for correct
installation.

1.4.2. COMMISSIONING

Follow the steps below to start up the booth:
NOTE: It is presupposed that the user has an audiometer available to connect
to the booth.
• Connect the mains power cable to the mains power supply input at the base of
the booth, and then connect it to the mains power supply.
• Check that turning on the main switch (Nº3) lights up the interior of the booth.
(If this is not the case, check the previous step and the mains power supply).
Leave it switched off.
• Place the audiometer on the tray and connect its mains power supply cable to
the output mains base (No. 3). IMPORTANT NOTE: Maximum load 500 W.
• Connect the audiometer accessories to the internal connections of the booth,
and hang them on the supports available.
• Remember the connection number for each accessory to later connect the
audiometer to the external connections of the booth (No.3), so that each
accessory is correctly connected to the audiometer.