10 S pecifications
S220 pH meter
Measurement range
Resolution
Limits of error pH
Limits of error ion
pH calibration
Isopotential point
pH Calibration buffer
Outputs
Power requirements
Size/weight
Display
pH input
Temperature input
Digital sensor input
Ambient conditions
Materials
44 Specifications
pH
mV
pH ATC
pH MTC
Ion
pH
mV
pH Temperature
Ion
± 0.002 pH
± 0.2 mV
± 0.1°C
± 0.5% (this limit only applies
for meter)
Up to 5 points
pH 7.00
7 predefined groups
RS232, USB A, USB B
DC9-12V-10W
204 x 174 x 74 mm
890 g
TFT
BNC,
impedance >
3 * 10e+12 Ω
RCA (Cinch), NTC 30kΩ,
Pt1000
Mini-LTW
Temperature
Relative humidity
Installation category
Pollution degree
Altitude
Housing
Window
Keypad
-2.000...20.000
-2000.0...2000.0 mV
-5...130°C
-30...130°C
0.000...1000.0%
0.000...10000 ppm
1.00E-9...9.99E+9 mg/L
1.00E-9...9.99E+9 mmol/L
1.00E-9...9.99E+9 mol/L
0.1/0.01/0.001
1/0.1
0.1°C
3 or 4 digits
1 user-defined group of 5 buf-
fers
5...40°C
5%...80% (non-condensing)
II
2
Up to 2000 m above sea level
ABS/PC reinforced
Polymethyl methacrylate
(PMMA)
Membrane keypad: Poly e the-
lene terephtalate (PET)