EN.
Display
Alternate (Blink)
Alternate (Blink)
*1 "Digit" is the minimum display unit.
For example, if minimum display is 0.1 g, 1 digit is equal to 0.1 g (50 digits are 5.0 g).
Error code
Sample mass error
The sample is too light to be stored as a sample mass for counting
mode or percent mode. Use a larger sample.
Unit mass error
The sample unit mass for counting mode is too light. Storing and
using it for counting will cause a counting error. Add samples to reach
the number indicated and press the PRINT key. Pressing the
PRINT key without adding samples will put the balance in counting
mode, but to ensure accurate counting, be sure to add samples.
Clock battery error
The clock backup battery has been depleted. Press any key and set
the time and date. Even if the clock backup battery is depleted, the
clock and calendar function works normally as long as the balance is
powered with the AC adapter. If this error appears frequently, contact
your local A&D dealer.
Power supply voltage fault
The voltage supplied from the AC adapter is abnormal. Make sure
you are using the AC adapter supplied with the balance.
ECL repeatability error
With the self-check function, the standard deviation (SD) of
repeatability with electronically controlled load (ECL) exceeded 50
digits
*1
. Check the installation environment of the balance.
ELC repeatability error
ELC minimum weighing value (reference value) error
Refer to "6-2. Self Check Function / Automatic Setting of
Minimum Weighing Value by ELC" in the separate manual.
Full memory
The number of stored weighing values has reached the upper limit. In
order to store new weighing values, it is necessary to delete data.
Refer to "11. Data Memory" in the separate manual.
Full memory
The number of items stored in your sensitivity adjustment / calibration
history has reached 50. When new items are added, older items are
deleted. Refer to "11. Data Memory" in the separate manual.
Level check
Check the bubble level on the main unit. Rotate the leveling feet to
center the bubble in the circle.
Either adjusting the bubble level or pressing any key other than the
ON:OFF key will clear the warning.
Description