2 Warning and safety messages
• Burn hazard! The areas in and around the sintering
chamber may be hot, especially on top (> 65°C). Wear
protective gloves.
• Burn hazard! Sintering trays may be hot. Wear protec-
tive gloves.
• Fire
Never
hazard!
Cercon
heat plus P8, not even temporarily.
®
• Do not reach into the sintering chamber. Burn hazard!
• Do not reach into the sintering chamber to prevent dam-
age to the retort tube or heating element.
• Crushing hazard when the lifting table closes! Keep away
from all movable parts.
• Always place the sintering trays in the middle of the in-
sulated floor of the sintering chamber. Observe the MAX
mark when stacking trays to avoid damage to the installa-
tion, the retort tube, the heating elements, or the thermal
element.
• Use only Cercon
heat plus sintering trays in the furnace.
®
Cercon heat sintering trays may not be used. Warning:
Improper tray use may result in damage to the heating
elements!
• Opening the unit, is prohibited. Electric shock hazard!
• If the Cercon
heat plus P8 is not to be used for an ex-
®
tended period, disconnect the power cord.
• Pull the power supply plug before use the emergency
lowering!
2.3 Maintenance and troubleshooting safety notes
Observe the following rules during maintenance, repairs,
or troubleshooting tasks:
• All repairs must be performed by authorized DeguDent
GmbH personnel only.
• Only qualified electricians are allowed to open the unit and
to work on electrical components.
• Disconnect the power cord before opening the unit.
• Do not insert any objects into the ventilation slits. Protect
from moisture.
• Power cords may only replaced by approved power cords
with an identical rating.
• Use only original replacement parts.
• The heating elements of Cercon
replaced with daily usage after 12 months.
30
place
any
objects
on
heat plus P8 are to be
®
2.4 Emergency shutdown
An emergency shutdown is performed by disconnecting
the power cord. During operation of the unit, the power
cord must be accessible at all times be allow disconnecting
the unit in an emergency.
the
2.5 Safe handling of ceramic fibers
The heat insulation of the sintering chamber consists of
ceramic fibers. Extended use or higher temperatures will
result in the formation of substances that, when released
into the air, may create a dust hazard, causing irritation
to the skin or eyes or the respiratory tract. The insula-
tion material used for this furnace contains ceramic fibres.
In the European Union, ceramic fibres have been rated as
follows, pursuant to EC Directive 97/69/EC dated
5 December 1997:
CARC.CAT 2; R49; Xi R 38.
Never open and expose the heat insulation of the
sintering chamber. Any repairs in this area must be
performed by authorized DeguDent GmbH personnel only.