EN-10
3.2
Menu Navigation
TO ENTER THE MENU MODE
Press and hold the Menu button until MENU appears on the display. The first upper level menu appears on
the display.
Summary of button navigation functions in menu mode:
--Yes
Allows entry into the displayed menu.
- Accepts the displayed setting and advances to the next menu item.
--No
Skips by the displayed menu.
- Rejects the displayed setting or menu item and advances to the next available item.
--Back
Moves backwards through the upper and middle level menus.
- Backs out of a list of selectable items to the previous middle level menu.
--Exit
Exits from menu directly to the active weighing mode.
3.3
Calibration Menu
Three calibration processes are available: Zero Calibration, Span Calibration and Linearity Calibration.
NOTES:
1. Make sure that appropriate calibration masses are available
before beginning calibration.
2. Make sure that the scale base is level and stable during the
entire calibration process.
3. Calibration is unavailable with LFT set to On.
4. Allow the Indicator to warm up for approximately 5 minutes
after stabilizing to room temperature.
5. To abort calibration, press the Exit button anytime during the
calibration process.
6. Make sure to finish Zero Calibration before performing Span
Calibration to ensure weighing accuracy.
3.3.1 Zero Calibration
Zero Calibration uses one point. The calibration point is established with no load on the scale. Use this
calibration method to adjust for a different pre-load without affecting the span or linearity calibration.
When [CAL] is displayed, press the YES key to accept the Calibration sub-menu selection.
(Pressing the NO key to advance to the next sub-menu, [SEtUP].)
When [ZErO] is displayed, press the YES key to accept the Zero Calibration menu item
selection.
The display flashes [0] and the kg led is light.
Press the Yes key to establish the zero point.
Note: The new zero point must be within the range of the normalized weight.
The display shows [--C--] while the zero point is established.
If zero calibration was successful, the scale exits to the next Calibration menu and displays
[SPAN]
CATAPULT 5000 SERIES SCALES
Zero
Perform
Span
Perform
Linearity
Perform
Set 00...Set 19... Set 31
Geographic
Adjustment
End Calibration
Exit CALIBRATE menu