SET-UP
NOTE– Before operating the DeVilbiss iFill Personal Oxygen Station for the first time, locate the device
where desired.
CHOOSING A LOCATION
The personal oxygen station should be at least 6 inches (15 cm) away from walls, draperies, or
Avoid placing the oxygen station next to heaters, radiators, or hot air registers. It should be kept
at least 5 feet (1.6m) away from hot, sparking objects or naked sources of flame.
Do not use in an area where the air may be contaminated with carbon monoxide as this may
shorten the life of the unit (i.e. near running gasoline engines, furnace, or heater).
TRANSPORTING THE DEVILBISS IFILL PERSONAL OxYGEN STATION
NEVER transport the Personal Oxygen Station with a cylinder connected
to or stored on it, otherwise injury or damage can occur.
CAUTION– When transporting the DeVilbiss iFill Personal Oxygen Station, be careful not to bump
the unit or the connectors into obstacles. Otherwise damage to the oxygen station may
occur.
There are three (3) methods to safely transport the DeVilbiss iFill Personal Oxygen Station
(Figure 1):
1. Pull or push the unit on its casters using the front handle. Travel in a line oriented front to
back to avoid tipping. Stop the unit to change direction.
2. Carry the unit using the front and rear handles simultaneously.
3. Carry the unit using the front handle and the recess located under the rear of the base.
DEVILBISS IFILL PERSONAL OxYGEN STATION OPERATION
CHECKLIST
NOTE– Each time the Oxygen Station is used to fill a cylinder, complete the following steps:
1. Check the Hydrostatic Test Date Figure C, 3 - Aluminum cylinders must undergo testing
every five (5) years. Contact your homecare provider or the cylinder manufacturer for details.
NOTE– Do not fill oxygen cylinders that have not been tested in the past five (5) years. Contact your
Sunrise provider for replacement.
2. Perform the prefill inspection on the cylinder by following iFill Oxygen Cylinder External
Examination below.
ONLY use cylinders that have the DeVilbiss iFill Personal Oxygen Station
All cylinders must be inspected before attempting to fill - otherwise, injury
IFILL OxYGEN CYLINDER ExTERNAL ExAMINATION
1. Examine the outside of the cylinder for the following conditions, and replace the cylinder if
they exist:
a) Dents or dings
b) Arc Burns
c) Oil or Grease
d) Any other signs of damage that might cause a cylinder to be unacceptable or unsafe for use.
2. Examine the cylinder for evidence of fire or thermal damage. Evidence includes charring or
blistering of the paint, or other protective coating or heat sensitive indicator. If fire or thermal
damage is found, replace the cylinder.
WARNING
furniture to ensure sufficient airflow.
Do not use in a closet.
WARNING
WARNING
fill connection.
or damage may occur.
11
H E A D E R
SE T- U P
Figure 1
SE-535