3. Inspect the oxygen fill connector for the following:
a) Debris, oil or grease
b) Noticeable signs of damage
c) Signs of corrosion inside the valve
d) Signs of excessive heat or fire damage
Do not use oxygen cylinder if found, call your homecare provider.
OPERATION
CONNECTING THE IFILL OxYGEN CYLINDER TO THE DEVILBISS
IFILL PERSONAL OxYGEN STATION
Fire Hazard. Do not lubricate. Do not allow grease or oil from your hands or other
source to come into contact with the regulator or cylinder valve connection. These con-
taminants may be flammable and cause injury.
Do not modify ANY connections on the personal oxygen station. NEVER
use tools of any kind to connect/disconnect the cylinder and the oxygen station.
Severe injury and/or damage may occur.
Do not drop oxygen cylinders. Use two (2) hands when handling/transporting
oxygen cylinders. Otherwise, injury or damage may occur.
NOTE– Refer to the General Dangers & Warnings and the Handling Warnings in this
manual.
1. Plug the power cord into the wall outlet.
2. Set the rotary selector on the cylinder to "OFF. "
3. Remove the oxygen fill connector cover from the fill connector.
4. Position the cylinder over the cradle while aligning the nipple connector on the oxy-
gen cylinder with the fill connector. Press until cylinder "Clicks" into place (Figure 2).
NOTE– When using the smaller M4 or M6 size cylinders, it is necessary to place the
oxygen fill connector cover/cylinder adapter in the cradle to help support the cylinder.
When using the larger ML6, C, D, or E cylinders, the oxygen fill connector cover/cylin-
der adapter is not necessary and can be allowed to hang over the side.
FILLING THE IFILL OxYGEN CYLINDER
NOTE– Do not use cylinder while filling.
NOTE– Refer to the General Dangers & Warnings and the Handling Warnings in this
manual.
1. Make sure the personal oxygen station is plugged in.
2. Make sure the cylinder is securely connected to the oxygen station. Refer to
Connecting The iFill Oxygen Cylinder To The DeVilbiss iFill Personal Oxygen Station.
3. Push the start/stop switch on the control panel (Figure 3).
4. The following sequence of events should occur:
a. The FILLING (green) light will illuminate while the cylinder is filling (refer to iFill
Oxygen Cylinder Fill Times).
b. The FULL (green) light will illuminate when the cylinder is full. Proceed to
Removing the iFill Oxygen Cylinder.
NOTE– If the Service (RED) light and audible alert are on, proceed to Trouble shooting in
this instruction manual.
SE-535
DANGER
WARNING
12
O P E R AT IO N
H E A D E R
Figure 2
Figure 3