Test bench MS006
- Insert the USB flash disk into the USB port of the test bench.
- Turn on the test bench. When started, the program will automatically detect the new version
and launch the installation.
- Wait until the installation is finished.
WARNING!
Do not interrupt the firmware update process. Turning the test bench off or
removing the USB flash disk when the installation is in progress is prohibited.
- After finishing the installation, the calibration window is downloaded. (Fig.28). Keep tapping the
touch pointer until the calibration is finished and the main menu is loaded.
- Turn off the test bench, remove the USB flash disk, and wait for 10 seconds. Now turn on the test
bench and use it as normal.
Figure 28. Touch screen calibration window
The update procedure for the test bench databases is as below:
- Download the new program version under the product page on servicems.eu.
- Copy folder "MS006Base" to the root directory of the USB flash drive.
WARNING!
The USB flash drive should not contain more than one "MS006Base" folder.
- Turn the test bench off.
- Insert the USB flash drive in the USB port of the test bench.
- Turn on the test bench. After the launch, the program will automatically find the new firmware
version and start the installation.
- Wait until the installation is finished. It may take some time.
WARNING!
Not to interrupt the installation please, do not turn the test bench off or remove
the USB flash drive.