- When the installation is finished, the main menu will be loaded. Now the test bench is ready for
operation.
In case of the touch screen improper function, calibrate it as below:
1. Turn the test bench off.
2. Press and hold down the "ELECTRICAL LOAD" knob.
3. Press "ON" to turn on the bench.
4. Press and hold down the "ELECTRICAL LOAD" knob until the touch screen calibration window is
loaded (Fig. 28).
5. Keep tapping the touch pointer until the calibration is finished and the main menu loaded.
6. Now the touch screen calibration is completed, and the test bench is ready for operation.
Use soft tissues or wipe cloths to clean the surface of the device with neutral detergents. Clean
the display with a special fiber cloth and a cleaning spray for touch screens. To prevent corrosion,
failure, or damage to the test bench, do not use any abrasives or solvents.
Table with the possible problems and the solutions on their elimination:
Failure symptom
1. The test bench does
not start.
2. Short-circuit sound
alert (bleep) is on when
turning on the test
bench.
3. The test bench is
running, but the electric
motor does not start.
7.2. Touch screen calibration
7.3. Cleaning and care
8. TROUBLESHOOTING GUIDE
Potential cause
The EMERGENCY STOP button is
on.
The supply voltage is under
220V.
Crocodile clips (+)/(-) closure to
the test bench body.
The electric wiring of the bench
is damaged.
Frequency converter failure.
The electric wiring of the bench
is damaged.
User manual
Troubleshooting tips
Turn off the emergency button.
Restore the supply voltage.
Break the clips.
Contact the sales
representative.
Contact the sales
representative.