6 Error limits
20 °C
!
Ex
Error limits
Nominal volume ml
10
*A = Accuracy, CV = Coefficient of Variation
V
N
.
A
=
T
V
T
e.g.
V
N
V
= 50% N
T
V
= 10% N
T
*A = Accuracy, CV = Coefficient of Variation
The error limits in DIN EN ISO 8655- 5 are significantly lower. The maximum error for a single measure-
ment can be approximated from the sum of error limits EL = A + 2 × CV (for 10 ml: 50 µl + 2 x 10 µl = 70
µl).
997491
Error limits in relation to the nominal capacity (= maximum
volume) indicated on the instrument at equal temperature
(20 ºC/68 °F) of instrument, ambient environment and dis-
tilled water. Testing takes place according DIN EN ISO 8655-
6 with a completely filled instrument and with uniform and
smooth dispensing.
A*
≤
0.5
Partial volumes
The percentage values for A and CV are relative to the nomi-
A
N
nal volume (V
(V
).
p
Volume
A*
≤
10.0
0.5
5.0
1.0
1.0
5.0
μl
± %
50
) and must be converted for partial volumes
N
μl
± %
50
50
50
NOTICE
6 Error limits
μl
CV*
≤
%
0.1
10
μl
CV*
%
≤
0.1
10
0.2
10
1.0
10
Operating manual