Signals and warnings
• Electrical shock risk: also 230 V
• Biological danger, risk of infections from epidemic diseases.
• General danger sign.
• High temperature.
• Compulsory direction of fl ow or of rotation.
ASPIRATION SYSTEM EQUIPPED WITH TURBO-SMART
ROUTINE MAINTENANCE CARD
Ordinary maintenance must be entrusted to specially instructed
and trained staff.
Maintenance date .................................... User signature ....................................
Every day: check the possible alarm signal on the display*
Every day: evening aspiration of the solution made up of Puli-Jet plus Disinfectant,
hot water and Pulse Cleaner
Occasionally: check the aspirator noisiness*
Disconnect the machine before any maintenance
Every day: replace the dirty fi lters with clean and disinfected fi lters, place correctly
the antifoaming disinfectant tablets on the fi lters of the dentist's unit
At every maintenance: make sure that the aspirator ventilation is not obstructed
(by dust, paper or other things)
Keep the plant room free from anything not concerning the machines, especially
from fl ammable material.
Make sure that there is no possibility for the formation of corrosive, fl ammable and
explosive mixtures.
* Contact the engineer in case of alarms or increased noisiness.
This maintenance sheet is a useful reference, but must be used in conjunction with the full user manual
On the website www.cattani.it it is possible to consult our updated manuals. We recommend the consultation especially
for updates about safety.
can be lethal.