Descargar Imprimir esta página

Hach 8310 Manual Del Usuario página 14

Ocultar thumbs Ver también para 8310:

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 42
Option
Description
Hold
The sensor output value is held at the current measured value during the calibration procedure.
Transfer A preset output value is sent during calibration. Refer to the controller user manual to change the
preset value.
6. Hold the dry sensor in the air and push OK.
7. Do not push OK until the calibration result shows on the display.
8. Review the calibration result:
• "The calibration was successfully completed."—The sensor is calibrated and ready to measure
samples. The slope and/or offset values are shown.
• "The calibration failed." —The calibration slope or offset is outside of accepted limits. Repeat
the calibration with fresh reference solutions. Clean the sensor if necessary.
9. Push OK.
10. Proceed to the calibration with a process sample.
4.3.7 Calibration with the process sample
Calibrate the sensor while the sensor is installed in the process sample. As an alternative, put the
sensor in a grab sample collected from the process sample.
Note: Before the sensor is calibrated for the first time, complete a zero calibration.
1. Select the main menu icon, then select Devices. A list of all of the available devices shows.
2. Select the sensor and select Device menu > Calibration.
3. Select Calibration again.
Note: Use the Measurement type setting to change the parameter that is calibrated.
4. Select the option for the output signal during calibration:
Option
Description
Active
The instrument sends the current measured output value during the calibration procedure.
Hold
The sensor output value is held at the current measured value during the calibration procedure.
Transfer A preset output value is sent during calibration. Refer to the controller user manual to change the
preset value.
5. With the sensor in the process sample, push OK.
The measured value is shown.
6. Wait for the value to stabilize and push OK.
Note: The screen may advance to the next step automatically.
7. Measure the conductivity (or other parameter) value with a secondary verification instrument. Use
the arrow keys to enter the measured value and push OK.
8. Review the calibration result:
• "The calibration was successfully completed."—The sensor is calibrated and ready to measure
samples. The slope and/or offset values are shown.
• "The calibration failed." —The calibration slope or offset is outside of accepted limits. Repeat
the calibration with fresh reference solutions. Clean the sensor if necessary.
9. Push OK to continue.
10. Return the sensor to the process and push OK.
The output signal returns to the active state and the measured sample value is shown on the
measurement screen.
4.3.8 Resistor calibration
Do a resistor calibration. A resistor calibration is an electrical calibration.
14 English

Publicidad

loading

Este manual también es adecuado para:

831183128315831683178394