Descargar Imprimir esta página

Hach 8310 Manual Del Usuario página 20

Ocultar thumbs Ver también para 8310:

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 42
Warning
Measurement is too high. The measured value is > 2 S/cm,
Measurement is too low.
Zero value is too high.
Zero value is too low.
Temperature is too high.
Temperature is too low.
Calibration is overdue.
The device is not
calibrated.
Replace a sensor.
Calibration is in
progress.
Outputs on hold
Linear temperature
compensation is out of
range.
Temperature
compensation table is
out of range.
Incorrect user
concentration table.
Incorrect built-in
temperature table.
Incorrect built-in
concentration table.
Section 8 Consumables and accessories
Consumables
Description
Conductivity standard solution, 25 µS/cm
Conductivity standard solution, 100 µS/cm
Conductivity standard solution, 12.88 mS/cm
20 English
Table 8 Warning list
Description
1,000,000 ppm, 200% or 20,000 ppt.
The measured value is < 0 μS/cm,
0 ppm, 0% or 0 ppt.
The zero calibration value is too high.
The zero calibration value is too low.
The measured temperature is >
200 °C.
The measured temperature is <
-20 °C.
The Calibration reminder time has
expired.
The sensor has not been calibrated.
The Sensor days counter is more
than the interval selected for sensor
replacement. Refer to .
A calibration was started but not
completed.
During calibration, the controller
outputs were set to hold for a
selected time.
The user-defined linear temperature
compensation is out of range.
The user-defined temperature
compensation table is out of range.
The concentration measurement is
outside of the range of the user table.
The measured temperature is outside
of the range of the built-in
temperature compensation table.
The concentration measurement is
outside of the range of the built-in
concentration table.
Resolution
Make sure that the display format is set for
the correct measurement range.
Make sure that the sensor is configured for
the correct cell constant.
Make sure that the sensor is held in air during
zero calibration and is not located near radio
frequency or electromagnetic interference.
Make sure that the cable is shielded by metal
conduit.
Make sure that the sensor is configured for
the correct temperature element.
Calibrate the sensor.
Calibrate the sensor.
Replace the sensor. Reset the Sensor days
counter on the Diagnostics/Test > Reset
menu (or Diagnostics/Test > Counter menu.
Return to calibration.
The outputs will become active after the
selected time period.
The value must be between 0 and 4%/°C;
0 to 200 °C.
The temperature is above or below the
temperature range defined by the table.
Make sure that the user table is set for the
correct measurement range.
Make sure that the temperature
compensation is configured correctly.
Make sure that the concentration
measurement is configured for the correct
chemical and range.
Quantity
500 mL
50 mL
500 mL
Item no.
S51M001
2971826
C20C250

Publicidad

loading

Este manual también es adecuado para:

831183128315831683178394