Figure 4 Connect the sensor cable
3.2 Connect the sensor cable to an SC Controller
1. Install an ultrapure conductivity module in the SC Controller. Refer to the installation instructions
supplied with the ultrapure conductivity module. Refer to
on page 20 for ordering information.
2. Connect the wires of the conductivity sensor cable to the ultrapure conductivity module. The
ultrapure conductivity module converts the analog signal from the sensor to a digital signal.
3.3 Mounting
Explosion hazard. Make sure that the mounting hardware for the sensor has a temperature and
pressure rating sufficient for the mounting location.
Install the electrode in the sample line or in a flow chamber. Refer to
on page 10 and
Figure 7
In the figures, the symbols A, B and C identify:
8 English
W A R N I N G
on page 10.
Consumables and accessories
Figure 5
on page 9,
Figure 6