Site preparation checklist
Safety requirements
Do those who will operate the printer have the technical training and
experience necessary to be aware of hazards to which they may be
exposed in performing a task, and to take appropriate measures to
minimize the risks?
Is there an emergency exit in the print production area, with easy access
and free from any obstruction?
Electrical installation requirements
Is the electrician aware of all requirements and specifications highlighted in
this guide?
Is the single-phase line voltage inside the specified voltage range (220–240
V~ (–10%, +6%))?
The tri-phase 208 V or 200 V power systems (voltage line-to-line) require a
step-up transformer.
Are there two dedicated lines to connect the printer's two power cords?
NOTE:
Do not use a power strip (relocatable power tap) to connect both
power cords.
Have branch circuit breakers (2 poles, 16 A general or 20 A for NEMA
6-20R only) been correctly installed for each dedicated line?
Has the single-phase branch circuit breaker (2 poles, 20 A) been correctly
installed?
Is the Power Distribution Unit (PDU) correctly installed?
Are the grounding conductors properly installed for each wall receptacle
(wall socket)?
Are the wall receptacles (wall sockets) suitable for the power cord plug type
provided by HP?
Are the wall receptacles (wall sockets) and electrical installation suitable for
the printer's rated current ?
NOTE:
The printer's rated current is 15 A.
8
Chapter 3 Site preparation checklist
Yes
No
Comments
(Required)
Yes
No
Comments
(Required)
(Required)
Specify nominal mains voltage:
(Required)
(Required)
(Required)
(Required)
(Required)
(Required)
(Required)