Possible option: Boiling in distilled
water
• Place the individual parts of your
nebuliser handset (incl. aerosol head) in
boiling water for at least 5 minutes. Use
a clean pot and distilled water for this.
CAUTION:
Plastic will melt if it comes in contact with
the hot bottom of the pot. Therefore, make
sure that there is sufficient water in the
pot. This will prevent the parts from being
damaged.
• Take the parts out of the pot.
• Shake all parts to remove excess water
more quickly.
• Place all nebuliser handset parts on a
dry, clean and surface and allow to dry
completely.
CAUTION:
Moisture promotes the growth of bacte-
ria. Please ensure adequate drying after
every disinfection.
CAUTION:
Check the nebuliser handset parts after
every disinfection. Replace all broken,
deformed and heavily discoloured parts.
®
eBase
Controller - 2022-09
Hygienic reprocessing in
the hospital and doctor's
practice
Carry out cleaning with disinfection
immediately after use. Ideally, a mechani-
cal procedure (washer-disinfector) should
be used.
Prepare for cleaning with disinfection (see
"CLEANING AND DISINFECTION" on
page 16).
Cleaning with thermal disinfection
• Place the nebuliser handset compo-
nents in the washer-disinfector.
• Select the 93 °C programme (treatment
time: 10 minutes).
The effectiveness of this procedure has
been validated using a G7736 disinfector
made by Miele and using neodisher MA
(Chemische Fabrik Dr. Weigert, Ham-
burg) as a cleaning agent in conjunction
with deionised water as a neutraliser.
Even when the washer-disinfector is
equipped with a drying function, ensure
that no residual moisture remains in the
nebuliser handset. Remove residual mois-
ture by shaking the parts and allow them
to air-dry completely.
en
19