Fault/Condition
The nebulisation time has become
noticeably longer.
Aerosol escapes continuously in
large volumes from the opening
slots in the medication reservoir.
The medication cap cannot be
removed from the medication
reservoir after the inhalation
treatment.
If the malfunction persists, contact your service partner.
®
eBase
Controller - 2022-09
Possible cause/Solution
If the inhalation time with the same fill
volume and the same substance becomes
considerably longer.
- You are inhaling a different medication.
- The fill volume has been increased.
- The medication cap was not closed
tightly.
- The aerosol head was not cleaned
immediately after the last application
(the fine pores of the membrane may be
partially blocked), or it is damaged.
- You are using the aerosol head for
longer than its recommended lifetime.
• Check whether the nebuliser handset
has been assembled correctly.
When doing so, check that the two valve
flaps of the inspiratory valve are seated
correctly. They must lie flat and face
towards the aerosol chamber.
The negative pressure in the medication
reservoir is too great.
• Open the aerosol chamber and gently
pull the aerosol head so that air is able to
flow in.
The medication cap can now be
removed easily.