TROUBLESHOOTING
6
You can correct malfunctions that occur during daily operation yourself. The following
tables will help you to locate and correct the causes of malfunction.
Screen
(flashes alternately)
®
eBase
Controller - 2022-09
Possible cause
Connection between
the controller and
the nebuliser
handset/aerosol
head interrupted.
No medication
added.
Inhalation
interrupted.
The maximum
operating time of
20 minutes for an
inhalation session
has been exceeded.
Battery charge
level 50%.
Battery flat.
Solution
- Verify that all parts
are securely
connected.
- Replace the
aerosol head and/
or the connection
cord if damaged.
If the fault persists,
contact your service
partner.
Add medication.
Resume inhalation
by restarting the
device (press
ON/OFF button).
Resume inhalation
by restarting the
device (press
ON/OFF button).
Have new batteries
on hand.
Insert new batteries
or operate with the
power adapter
connected.
en
21