Fault/Condition
The controller cannot be activated
(no beep, no orange or green light
signal).
After switching the controller on, no
aerosol is generated, and/or the
controller switches off again after a
few seconds.
The controller switches off, even
though there is still medication in the
medication reservoir.
Info:
Note for the eFlow
nebuliser handset that about
1 ml of medication cannot be
nebulised and remains unused
in the medication reservoir; this
residue must be discarded.
This is intentional and does not
represent a malfunction.If the
®
Tolero
nebuliser handset is
used no significant residual
volume remains in the
medication reservoir.
The controller does not switch off
automatically even when no more
aerosol is being produced and there
is no more than about 1 ml medica-
tion left in the medication reservoir if
®
a eFlow
rapid nebuliser handset is
used.
22
Possible cause/Solution
- To switch on, press the ON/OFF button
- Is power cord plug plugged into mains
- Are batteries or rechargeable batteries
- Check the charge level of the batteries.
- Has the medication reservoir been filled
- Check the connection between the
If there is still more than 1 ml in the
medication reservoir of the eFlow
nebuliser handset:
- Check the charge level of the batteries
®
rapid
- The nebuliser handset was not being
- The controller switches off automatically
The treatment session has been
completed successfully and the controller
can be switched off by pressing the
ON/OFF button.
for approx. 2 seconds.
and the device (for mains operation)?
inserted correctly?
with medication?
controller and the nebuliser handset.
or rechargeable batteries!
held horizontally. Hold the nebuliser
handset horizontally.
after 20 minutes.
Carry out inhalation to the end by
pressing the ON/OFF button again, so
that you can inhale the prescribed dose
of medication.
If the maximum operating time for each
inhalation session is exceeded repeat-
edly, the aerosol head must be replaced.
eBase
®
rapid
®
Controller - 2022-09