11.6 Settings Screen
11.6.1 Setting Units
The unit of measure for temperature can be selected in the Settings Menu. The
options available for temperature are °C/°F.
11.6.2 Choosing the Coolant
The coolant (Water/ Ethylene Glycol) can be chosen on the Settings Menu. The
Ethylene Glycol Percentage can be chosen once the EGW button is pressed. This
selection limits the temperature set point value that the customer can select.
Below are the temperature range for different glycol percentages. Note that
selecting the correct coolant is the responsibility of the customer and should
match what is filled in the system by the customer. Selecting the wrong coolant
may cause damage to the equipment.
Water: 10°C to 40°C
10% Ethylene Glycol: 5°C to 40°C
20% Ethylene Glycol: 0°C to 40°C C