Optional Automatic Wash / Disinfect
CAUTIONS
• Alkaline and highly acidic detergents and solutions cause damage to the devices.
• Channels and crevices found on these devices require particular attention during cleaning.
• Pay special attention to the water quality used throughout reprocessing. Hard water can damage the surface of the
equipment. Avoid using hard water. Instead use purified water unless otherwise specified.
Limitations on Reprocessing
• Repeated processing has negative effects on these devices and is not recommended for routine use.
• It is important to have Integra NeuroSpecialists perform routine inspections (twice yearly is recommended).
See contact information.
INSTRUCTIONS
Cleaning - Automated Equipment: Neutral pH Detergent
Method
1. Prepare equipment for cleaning by disassembling removable parts and loosening connections.
2. Rinse equipment in warm water before placing into washer.
3. Load device into the washer and place small parts in container or tray inside the washer unit in order to avoid losing
small components.
NOTE – Load devices carefully into washer in order to avoid collision.
4. Follow the instructions listed below and set washer machine to these exact parameters:
Phase
Time (Min.)
Pre-wash 1
4:00
Enzyme Wash
4:00
Wash 1
10:00
Rinse 1
0:30
Thermal Rinse**
2:00
** Optional phase for disinfection of components – minimum water temperature as indicated or per worker manufacturer
specifications for the thermal rinse cycle.
NOTE – Any deviation from this guideline could result in damage to the equipment as well as improper cleaning results.
Rinse with purified water. Do not perform if parameters cannot be achieved.
5. Remove from washer and dry completely if needed.
6. Inspect equipment to make sure there is no visible organic debris or residue from cleaning agent.
Repeat process if any visible soil is detected.
The life expectancy of the MAYFIELD products is expected to be as long as 7 years.
Water Temperature
Detergent and Concentration
Cold water
N/A
Hot water
Neutral pH enzymatic (e.g. Endozime® AW Triple Plus with
APA, Ratio 1:128)
60.0ºC (140º F)
Neutral pH detergent
(e.g. Renu-Klenz™ , Steris, Ratio 1:256)
Hot water
N/A
82.2ºC (180º F)
N/A
11
EN – ENGLISH